Campbell Scientific CNR1 Instruction Manual
![Campbell Scientific CNR1 Radiometer manual cover](/images/qXNVvKrldzGPUejgdrwlXJHE/thumbs/thumb175x175.webp)
To install the CNR1 Radiometer, mount it on a stable platform away from obstructions. Ensure the radiometer is level using the built-in bubble level, and securely fasten it to prevent movement. Connect the appropriate cables to your data logger and configure the logger to record readings as per the manual instructions.
Check all connections to ensure they are secure. Inspect the data logger settings to confirm it is configured to read the CNR1. Verify the power supply to the radiometer and data logger. If the issue persists, consult the troubleshooting section in the user manual.
Calibration of the CNR1 Radiometer should be performed by qualified personnel. It is recommended to return the unit to the manufacturer or an authorized calibration center. Calibration involves comparing the radiometer readings against a reference instrument under controlled conditions.
Regularly inspect the radiometer for any physical damage or dirt accumulation. Clean the sensor surfaces as needed, and check the desiccant status every six months. Replace the desiccant if it is saturated. Conduct a full performance check annually or as specified in the manual.
To clean the sensors, use a soft, damp cloth with mild detergent. Gently wipe the surfaces, avoiding excessive pressure. Do not use abrasive materials or solvents, as these can damage the sensor surfaces.
The CNR1 Radiometer is designed to operate in various weather conditions, including temperatures ranging from -40°C to 70°C. However, ensure the unit is not exposed to conditions outside its specified limits to avoid damage or inaccurate readings.
The CNR1 Radiometer provides outputs for net radiation, which includes both shortwave and longwave radiation components. These are typically measured in watts per square meter (W/m²). Ensure your data logger is set to capture these outputs correctly.
To replace the desiccant, carefully open the desiccant holder. Remove the old desiccant and replace it with fresh material. Ensure the holder is sealed properly to maintain sensor performance. Refer to the manual for detailed instructions.
On-site verification can be done by comparing the CNR1 readings with a calibrated reference radiometer under stable conditions. Ensure both instruments are positioned similarly and environmental conditions are stable during the comparison.
Common errors may include connectivity issues or out-of-range readings. Consult the user manual for specific error codes and corrective actions, such as checking connections, recalibrating the unit, or addressing environmental factors affecting measurements.